Measuring of excess air directly in the flue-gas after combustion.
Measuring of oxygen content in protective gasses, for instance
in connection with packaging of food products.
Facilitates regulation and thus minimum consumption of gases.
Measuring and control of the consumption of protective gasses
in connection with welding of for instance stainless steel.
- Power supply: 230 V AC 50/60 Hz
- Power comsumption: Approx. 30 VA
- Dimensions: H x W x D 244 x 157 x 65 mm
- Measuring range: 0.5 - 20.9 % O2 volume %
- Display: Seven segment L.E.D. display 0.5 - 20.9 % O2
- Accuracy: ± 0.15%
- Gas temperature: Maximum 600 °C
- Output signals: 4(0)-20 mA current loop signal
and two break contacts
- Galvanic isolation: 300 V DC (OC 2010 only)
- Alarms: 2 individually set levels 0.5 - 20.9 %
- Heat up time: Approx. 120 seconds
- Calibration: Atmospheric air and 2 % O2 in N2 once a year
- Dive length: 29 mm
- Dive length extr. tube: Standard: 460 mm (longer tubes on request)
- Probe connections: Minimum 2 meters of teflon cable
between probe and measuring amplifier
ending in a multiplug